- 1.1. If the equipment does not power on:
1. Check if the emergency button on the front is pressed; if pressed, turn it counterclockwise(CCW) to restore it to its original position.
2. Check the status of cord wires, circuit breakers, and fuses on the rear.
- a. If the cord wire is pulled, reconnect the power wire.
- b. If the breaker is off, turn it on.
- c. Replace the fuse if it is broken. - 1.2. If the lamp does not turn on:
1. Check if the lamp and lamp holder are properly installed.
2. Request A/S from us for transformer replacement.
(high voltage(HV) flowing to the lamp causing transformer failure.) - 1.3. If nanofiber collection is abnormal:
1. Verify that the collector is connected to ground(GND).
- a. Use a multimeter to measure the resistance between the collector and the ground(GND) of the power outlet or the ground(GND) of the multi-tap.
- b. With power, the resistance value must be close to zero(0 [Ω]) or zero (0 [Ω]).
- c. For drum collectors: first, measure the resistance between the collector (roll) and the ground(GND) end on the side with a multimeter.
- d. If there is resistance, the shaft bearing on the drum collector must be replaced.
- e. When the web thickens, apply (-) or (+) DC voltage to the collector to increase the potential difference.
(When the web thickens, the (+) or (-) charge generated by the high voltage(HV) generator does not leak into the collector, and the same charge may form on the surface of the membrane, preventing the electric field from forming.) - 1.4. If electrical noise occurs:
1. Electrical noise: instantaneous noise, such as "tyriric" and "tick-tick.“
2. Turn off the power immediately and contact us.
(There is a risk of fire.) - 1.5. If mechanical noise occurs:
- Mechanical noise: continuous noise, such as "giggling" and "crushing."
- Check if the cover part of the robot mechanism touches.
- Check if there is bearing wear noise in the LM block part.
(LM block part of the drive mechanism part needs to be replaced)